aalto1 untyped-item.component.html
Resilient flow control for wireless data streaming in inductively coupled medical implants
Loading...
Access rights
openAccess
publishedVersion
URL
Journal Title
Journal ISSN
Volume Title
A1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä
This publication is imported from Aalto University research portal.
View publication in the Research portal (opens in new window)
View/Open full text file from the Research portal (opens in new window)
View publication in the Research portal (opens in new window)
View/Open full text file from the Research portal (opens in new window)
Unless otherwise stated, all rights belong to the author. You may download, display and print this publication for Your own personal use. Commercial use is prohibited.
Date
Major/Subject
Mcode
Degree programme
Language
en
Pages
12
Series
Microprocessors and Microsystems, Volume 72
Abstract
This paper describes the hardware implementation of a custom communication protocol tailored for low power telemetry data streaming over an inductive link. An efficient transceiver design is achieved by adapting only the essential physical layer features of a typical RFID baseband processor and optimizing the flow control logic for continuous and reliable data transfer. For the external near-field reader, we provide a logical model for receiver operation and suggest a simple forward error correction (FEC) mechanism. The benefit of FEC in the context of developed communication system is demonstrated by simulations, and projections of design scalability are also presented. The proposed communication system was implemented in 28nm CMOS process. Place-and-route (PNR) results occupy only 0.0048 mm2 of core area, and the transient simulations show a power consumption of 306 nW at 0.5 V supply and a master clock of 845.7 kHz. The implementation provides an uplink rate of 12 kbit/s, sufficient for reliable transmission of a 1-channel 1 kS/s 12-bit sample recording.
Description
Other note
Citation
Kempi, I, Ahmed, N, Hammer, A, Olabode, O, Unnikrishnan, V, Kosunen, M & Ryynänen, J 2020, 'Resilient flow control for wireless data streaming in inductively coupled medical implants', Microprocessors and Microsystems, vol. 72, 102905. https://doi.org/10.1016/j.micpro.2019.102905